Show simple item record

Authordc.contributor.authorToral Ponce, María es_CL
Authordc.contributor.authorRichter Duk, Pablo 
Admission datedc.date.accessioned2011-06-15T13:07:31Z
Available datedc.date.available2011-06-15T13:07:31Z
Publication datedc.date.issued1996-05
Cita de ítemdc.identifier.citationMICROCHEMICAL JOURNAL 53 (4): 413-419es_CL
Identifierdc.identifier.issn0026-265X
Identifierdc.identifier.urihttps://repositorio.uchile.cl/handle/2250/121294
General notedc.descriptionArtículo de publicación ISIes_CL
Abstractdc.description.abstractAn integrated retention/photometric detection method for the determination of microamounts of iron was developed. The method is based on the retention of the reaction product of the analyte with ferrozine in a flow cell packed with an anion exchange support, and on the subsequent elution of the retained complex with a solution of 1.5% sodium peroxide, which allows the reuse of the system. The method is highly selective and has a detection limit (3 sigma) of 4.3 ng ml(-1). Iron was thus determined in the range 10-200 ng ml(-1) with a relative standard deviation of 1-3% (n = 11). The method was applied to the determination of iron in tap and river water and the results were consistent (2%) with those provided by AAS.es_CL
Lenguagedc.language.isoenes_CL
Publisherdc.publisherACADEMIC PRESS INC JNL-COMP SUBSCRIPTIONSes_CL
Keywordsdc.subjectRETENTION SPECTROPHOTOMETRIC DETECTIONes_CL
Títulodc.titleFlow injection with an integrated retention/photometric detection unit for the determination of iron in wateres_CL
Document typedc.typeArtículo de revista


Files in this item

Icon

This item appears in the following Collection(s)

Show simple item record