Validated HPLC method for the determination of ranitidine in plasma
Artículo
Open/ Download
Publication date
2003-10Metadata
Show full item record
Cómo citar
Castro, A.
Cómo citar
Validated HPLC method for the determination of ranitidine in plasma
Author
Abstract
A validated HPLC method for the determination of ranitidine in human plasma is presented. Sulfanilamide as internal standard (IS) was used. Plasma samples were purified by solid phase extraction (SPE) using a copolymeric [poly(divinyl-benzene-co-N-vinylpyrrolidone)] column ("Oasis Waters"). Mobile phase consisting of dibasic potasium phosphate 0.08 M/acetonitrile/methanol/triethylamine 0.05% (89.5:3:7:0.05) pH5 was used at a flow rate of 0.9 ml/min on a C18 column (Nova-Pack, 3,9 x 300 mm, Waters). The eluate was monitored using an UV/Vis detector set at 300 nm. Ratio of peak area of ranitidine to sulfanilamide was used for the quantitation of plasma samples. FDA criteria for bioanalytical validation was used to validate the method. Linearity was assessed between 100-1600 ng/ml, the limit of quantitation was 100 ng/ml and recovery was greater than 94%. Accuracy, precision and selectivity met the current recommendations for bioanalytical method validation. The method was successfully used in a bioavailability study of a ranitidine tablet in healthy volunteers.
Quote Item
PHARMAZIE 58 (10): 696-698