Carbon nanofiber screen printed electrode joined to a flow injectionsystem for nimodipine sensing
Artículo
Open/ Download
Publication date
2015Metadata
Show full item record
Cómo citar
Salgado Figueroa, Paola
Cómo citar
Carbon nanofiber screen printed electrode joined to a flow injectionsystem for nimodipine sensing
Abstract
tA simple and sensitive electrochemical method using pretreated carbon nanofiber screen printed elec-trodes joined to a flow injection analysis (CNF-SPE-FIA) system was developed to analyze nimodipine(NMD) in pharmaceutical formulations. Low detection limits were achieved by controlling the accumula-tion time or sample volume. The developed method performed well with a LOD from 0.08 to 0.400 M fora 6 or 1 mL sample loop and an RSD below 2% for a wide linear range. This method was applicable to bothnamed brand and generic drug tablets containing 30 mg NMD. A contents concentration of 30.2 mg/tabletof NMD with an RSD = 2.0% for named brand and 30.1 mg/tablet with an RSD = 1.9% of NMD in the genericdrug was determined. This work could promote the potential application of CNF-SPE-FIA system as a newstrategy to be applied to the analysis of any drug derived from nitroaromatic compounds.
General note
Artículo de publicación ISI
Patrocinador
AcknowledgementThis work was supported by Grant No. 1130160 from FONDECYTChile.
Identifier
URI: https://repositorio.uchile.cl/handle/2250/135510
Quote Item
Sensors and Actuators B 220 (2015) 456–462