The rupture of surgical gloves in a general hospital Rotura de guantes quirúrgicos en un hospital general.
Artículo
Open/ Download
Publication date
1990Metadata
Show full item record
Cómo citar
Salas,
Cómo citar
The rupture of surgical gloves in a general hospital Rotura de guantes quirúrgicos en un hospital general.
Author
Abstract
We studied the incidence of rupture of surgical gloves used at a general hospital: A) new gloves (n = 114); B) gloves used at a major general, gynecological or urological operation (n = 258), and C) recirculating surgical gloves (n = 86). A special techniques based on transmission of an electrical signal through the rupture was used to identify rupture. Rupture was present in 8.8, 28 and 26% of surgical gloves in groups A, B and C, respectively. Type of surgery or surgeon involved did not influence rupture of gloves in group A. However, rupture increased in relation to the duration of surgery (40% incidence in operations lasting more than 90 min). Our results suggest the need for better quality control of new surgical gloves and glove change after 90 min of operation.
Indexation
Artículo de publicación SCOPUS
Quote Item
Revista medica de Chile, Volumen 118, Issue 1, 2018, Pages 44-48