Atrioventricular plane displacement versus mitral and tricuspid annular plane systoli excursion: A comparison between cardiac magnetic resonance and M-mode echocardiography
Artículo
Open/ Download
Access note
Acceso abierto
Publication date
2021Metadata
Show full item record
Cómo citar
Sepúlveda Martínez, Álvaro Gabriel
Cómo citar
Atrioventricular plane displacement versus mitral and tricuspid annular plane systoli excursion: A comparison between cardiac magnetic resonance and M-mode echocardiography
Author
Abstract
Introduction: Both echocardiography and CMR imaging are used to quantify longitudinal
function. Inter-method
variability for mitral (MAPSE) and tricuspid (TAPSE)
annular plane systolic excursion, and variability between directly measured MAPSE
and TAPSE and as based on atrioventricular plane displacement (AVPD) analysis by
CMR, are, however, not known. This study, therefore, assessed inter-method
variability
and variability between annular plane systolic excursion and AVPD-based
values
in a healthy adult population.
Methods: Echocardiography and CMR were performed in 111 adults (35 [32–38]
years). Method comparisons were assessed with Deming regression, Bland–Altman
analysis and coefficient of variation. Observer reproducibility was assessed by the
concordance correlation coefficient.
Results: Echocardiography and semi-automatic
CMR agreed on MAPSE (17 ± 2 mm
vs. 17 ± 2 mm, p = 0.1) and TAPSE (25 ± 3 mm vs. 25 ± 3 mm, p = 0.5), correlated
highly between methods (fitted-slope
1.22 [95% CI 1.07–1.38]
and 1.12 [95% CI 0.95–1.29])
and showed low bias (0.42 [95% CI − 2.05 to 2.88] and − 0.18 [95% CI − 4.78
to 4.43]). Intra-/
inter-observer
reproducibility was high for both methods for both.
MAPSE (echocardiography 0.96/0.86; CMR 0.87/0.85) and TAPSE (echocardiography
0.96/0.95; CMR 0.97/0.96). MAPSE (16 ± 2 mm vs. 17 ± 2 mm; p < 0.001) and TAPSE
(24 ± 3 vs. 25 ± 3 mm; p < 0.001) based on AVPD were similar but statistically different
compared with semi-automatic
CMR.
Conclusions: Echocardiography and semi-automatic
CMR have low variability and
provide similar values for MAPSE and TAPSE and are thus interchangeable for follow-
up
studies. Lateral values based on tracked data from AVPD analysis are not clinically
significantly different and could be used as a representation of annular displacement.
Patrocinador
Erasmus Mundus
Region of Skane
Skane University Hospital Lund
Medical Faculty at Lund University
Knut & Alice Wallenberg Foundation
Instituto de Salud Carlos III
European Commission PI15/00130
PI14/00226
PI17/00675
INT16/00168
por el ISCIII-Subdireccion General de Evaluacion y el Fondo Europeo de Desarrollo Regional (FEDER) `Otra manera de hacer Europa' (Spain)
Erasmus + Programme of the European Union 2013-0040
La Caixa Foundation LCF/PR/GN14/10270005
AGAUR 2014 SGR grant 928
Swedish Heart-Lung Foundation
Indexation
Artículo de publícación WoS Artículo de publicación SCOPUS
Quote Item
Clinical Physiology and Functional Imaging (2021) 41:3 Págs. 262 - 270
Collections
The following license files are associated with this item: